Journal article

Diurnal Patterns in Solute Concentrations Measured with In Situ UV-Vis Sensors: Natural Fluctuations or Artefacts?


Authors listJacobs, SR; Weeser, B; Rufino, MC; Breuer, L

Publication year2020

Pages859-

JournalSensors

Volume number20

Issue number3

Open access statusGold

DOI Linkhttps://doi.org/10.3390/s20030859

PublisherMDPI


Abstract
In situ spectrophotometers measuring in the UV-visible spectrum are increasingly used to collect high-resolution data on stream water quality. This provides the opportunity to investigate short-term solute dynamics, including diurnal cycling. This study reports unusual changes in diurnal patterns observed when such sensors were deployed in four tropical headwater streams in Kenya. The analysis of a 5-year dataset revealed sensor-specific diurnal patterns in nitrate and dissolved organic carbon concentrations and different patterns measured by different sensors when installed at the same site. To verify these patterns, a second mobile sensor was installed at three sites for more than 3 weeks. Agreement between the measurements performed by these sensors was higher for dissolved organic carbon (r > 0.98) than for nitrate (r = 0.43-0.81) at all sites. Higher concentrations and larger amplitudes generally led to higher agreement between patterns measured by the two sensors. However, changing the position or level of shading of the mobile sensor resulted in inconsistent changes in the patterns. The results of this study show that diurnal patterns measured with UV-Vis spectrophotometers should be interpreted with caution. Further work is required to understand how these measurements are influenced by environmental conditions and sensor-specific properties.



Citation Styles

Harvard Citation styleJacobs, S., Weeser, B., Rufino, M. and Breuer, L. (2020) Diurnal Patterns in Solute Concentrations Measured with In Situ UV-Vis Sensors: Natural Fluctuations or Artefacts?, Sensors, 20(3), p. 859. https://doi.org/10.3390/s20030859

APA Citation styleJacobs, S., Weeser, B., Rufino, M., & Breuer, L. (2020). Diurnal Patterns in Solute Concentrations Measured with In Situ UV-Vis Sensors: Natural Fluctuations or Artefacts?. Sensors. 20(3), 859. https://doi.org/10.3390/s20030859


Last updated on 2025-10-06 at 11:09